Abstract
Objective: The objective of the study is to compare radiofrequency (RF) effects on fresh placentae with varying levels of sustained time (Ts) and degrees of target temperature (°t). Method: A total of 108 pieces of fresh placentae were coagulated with a 2-cm RF needle at 60 W in an organ bath. The vertical and horizontal diameters (Vd, Hd) of tissue coagulation visualized by ultrasound were measured. The impacts of 12 different Ts-°t combinations on the ablation size ascertained on pathological examination (Vdp, Hdp) were compared using 2-way ANOVA. The agreement between sonographic and pathological findings was assessed using Bland-Altman analysis. Results: Considerable changes in the Vdp and Hdp were associated with increasing the Ts and °t. The impact of RF on tissue coagulation was greatest when the °t was set at 100°C, with further destruction as the Ts progressed to 7 minutes of exposure. The ablation size estimated by ultrasound exhibited an overestimation by an average of 5.65% and 21.02% for Vd and Hd, respectively. Conclusion: A prolonged Ts at a higher °t contributes to progressive placental tissue destruction by RF, with maximum destruction at 100°C for 7 minutes in an ex vivo nonperfused placenta. Tissue injury that is apparent on ultrasound may extend beyond pathological damage.
| Original language | English |
|---|---|
| Pages (from-to) | 504-510 |
| Number of pages | 7 |
| Journal | Prenatal Diagnosis |
| Volume | 38 |
| Issue number | 7 |
| DOIs | |
| Publication status | Published - Jun 2018 |
Fingerprint
Dive into the research topics of 'Time and temperature dependence of radiofrequency ablation in the human placenta'. Together they form a unique fingerprint.Cite this
- APA
- Author
- BIBTEX
- Harvard
- Standard
- RIS
- Vancouver