Abstract
This work exploited the well-known iodine-starch reaction for development of a simple flow-injection (FI) method for determination of iodide in pharmaceutical samples. Iodide in an injected zone was oxidized to iodine. A gas diffusion unit enables selective permeation of iodine through a hydrophobic membrane. Detection was made very selective for elemental iodine by employing formation of the I3 --starch complex. The detection limit (3S/N) of the system was 1mgIL-1. For a liquid patent medicine used for asthma treatment we suggested modification of the system. Direct injection of this sample, which contains a particularly high concentration level of iodide (ca. 9000 mgIL-1), can be achieved by coupling a dialysis unit to the FI system. This has increased the working range to 6000-10,000 mgIL -1 without employing complicated nanoliter injection.
| Original language | English |
|---|---|
| Pages (from-to) | 816-821 |
| Number of pages | 6 |
| Journal | Analytical and Bioanalytical Chemistry |
| Volume | 378 |
| Issue number | 3 |
| DOIs | |
| Publication status | Published - Feb 2004 |
UN SDGs
This output contributes to the following UN Sustainable Development Goals (SDGs)
-
SDG 3 Good Health and Well-being
Keywords
- Dialysis
- Flow injection
- Gas diffusion
- Iodide
- Iodine-starch
Fingerprint