Skip to main navigation Skip to search Skip to main content

Real-time inverse planning for Gamma Knife radiosurgery

  • Q. Jackie Wu
  • , Vira Chankong
  • , Suradet Jitprapaikulsarn
  • , Barry W. Wessels
  • , Douglas B. Einstein
  • , Boonyanit Mathayomchan
  • , Timothy J. Kinsella
  • University Hospitals of Cleveland
  • Case Western Reserve University

Research output: Contribution to journalArticlepeer-review

31 Citations (Scopus)

Abstract

The challenges of real-time Gamma Knife™ inverse planning are the large number of variables involved and the unknown search space a priori. With limited collimator sizes, shots have to be heavily overlapped to form a smooth prescription isodose line that conforms to the irregular target shape. Such overlaps greatly influence the total number of shots per plan, making pre-determination of the total number of shots impractical. However, this total number of shots usually defines the search space, a pre-requisite for most of the optimization methods. Since each shot only covers part of the target, a collection of shots in different locations and various collimator sizes selected makes up the global dose distribution that conforms to the target. Hence, planning or placing these shots is a combinatorial optimization process that is computationally expensive by nature. We have previously developed a theory of shot placement and optimization based on skeletonization. The real-time inverse planning process, reported in this paper, is an expansion and the clinical implementation of this theory. The complete planning process consists of two steps. The first step is to determine an optimal number of shots including locations and sizes and to assign initial collimator size to each of the shots. The second step is to fine-tune the weights using a linear-programming technique. The objective function is to minimize the total dose to the target boundary (i.e., maximize the dose conformity). Results of an ellipsoid test target and ten clinical cases are presented. The clinical cases are also compared with physician's manual plans. The target coverage is more than 99% for manual plans and 97% for all the inverse plans. The RTOG PITV conformity indices for the manual plans are between 1.16 and 3.46, compared to 1.36 to 2.4 for the inverse plans. All the inverse plans are generated in less than 2 min, making real-time inverse planning a reality.

Original languageEnglish
Pages (from-to)2988-2995
Number of pages8
JournalMedical Physics
Volume30
Issue number11
DOIs
Publication statusPublished - Nov 2003
Externally publishedYes

Keywords

  • Gamma Knife
  • Inverse planning
  • Optimization
  • Radiosurgery

Fingerprint

Dive into the research topics of 'Real-time inverse planning for Gamma Knife radiosurgery'. Together they form a unique fingerprint.