Abstract
An amperometric L-glutamate oxidase (GLOD) electrode based on a multilayer of polymer films was developed for the high selective determination of L-glutamate. The multilayer film consisted of three layers as the following configuration i.e., inner membrane of electropolymerized 1,3-diaminobenzene, middle layer of L-GLOD entrapped in photopolymerized poly(vinylferrocene)-poly(ethylene glycol) hydrogel polymer, and outer dialysis membrane. In this manner, the sensor could eliminate interferences and was able to work at a low potential poised at +0.085 V vs. Ag/AgCl. When used in a flow injection system, the sensor responded to L-glutamate in the range 0.5-8.0 mM, with a sensitivity of 9.48 nA mM-1. The sensor was stable for 5 days of continuous uses (250 assays) and retained 60% activity after 16 days. When used to analyse the L-glutamate contents of a number of different off-the-shelf soy sauces, the sensor gave results in good agreement with the standard colorimetric method.
| Original language | English |
|---|---|
| Pages (from-to) | 479-485 |
| Number of pages | 7 |
| Journal | World Journal of Microbiology and Biotechnology |
| Volume | 19 |
| Issue number | 5 |
| DOIs | |
| Publication status | Published - Jul 2003 |
| Externally published | Yes |
Keywords
- Biosensor
- Flow injection analysis
- L-glutamate oxidase
- Mediator
- Polymerization
Fingerprint
Dive into the research topics of 'Poly(vinylferrocene)-poly(ethylene glycol) glutamate oxidase electrode for determination of L-glutamate in commercial soy sauces'. Together they form a unique fingerprint.Cite this