Performance of a new interference-resistant glucose meter

  • Faculty of Medicine, Thammasat University
  • VA Medical Center

Research output: Contribution to journal › Article › peer-review

31 Citations (Scopus)

Abstract

Objectives: Glucose meters are widely used in self and hospital monitoring of blood glucose. We examined the analytical performance of a StatStrip glucose monitoring system. Design and methods: Linearity, % recovery and within-run imprecision were studied using glucose-spiked whole blood. A total of 120 heparinized samples were used in method comparison using a plasma hexokinase on the Dimension RxL MAX analyzer as the comparison method. Common interferences were tested on the StatStrip, Accu-Chek Advantage and the MediSense Optium glucose meters at low, middle and high glucose levels. Results: The StatStrip assay showed excellent linearity and recovery. The coefficient of variations for imprecision were < 5%. This meter correlated well with the comparison method (y = 0.994X + 0.03; r = 0.995, Sy/x = 0.05 mmol/L, bias = - 0.01 mmol/L). Of the three meters tested, only the StatStrip showed interference < 10% for all spiked levels of acetaminophen, ascorbic acid, maltose and hematocrit at three levels of glucose tested. Conclusions: The StatStrip meter showed good performance and is suitable for point-of-care hospital glucose testing.

Original languageEnglish
Pages (from-to)186-192
Number of pages7
JournalClinical Biochemistry
Volume43
Issue number1-2
DOIs
Publication statusPublished - Jan 2010
Externally publishedYes

Keywords

  • Glucose meter
  • Interference-resistant glucose meter
  • POCT
  • Point-of-Care-Testing
  • Self-monitoring of blood glucose (SMBG)

Fingerprint

Dive into the research topics of 'Performance of a new interference-resistant glucose meter'. Together they form a unique fingerprint.

Cite this