Skip to main navigation Skip to search Skip to main content

Non-Fourier encoding with multiple spin echoes

  • Brigham and Women's Hospital
  • Boston University College of Engineering

Research output: Contribution to journal › Article › peer-review

21 Citations (Scopus)

Abstract

The advantages and limitations of multiple spin-echo sequences for non- Fourier encoding are investigated. Complications caused by improper encoding of alternate magnetization pathways due to imperfect refocusing pulses are analyzed. It is shown that mirror image ghosts result if the encoding RF pulse matrix is real-valued. These ghosts can be avoided as long as the rows of the RF pulse matrix are conjugate symmetric, which implies that spatial profiles are real valued. Non-Fourier encoding using bases derived from wavelet, Hadamard, and other real-valued orthogonal functions does not result in a mirror ghost artifact. A RARE sequence for non-Fourier encoding has been implemented on a clinical imaging system and successfully applied for brain imaging.

Original languageEnglish
Pages (from-to)964-973
Number of pages10
JournalMagnetic Resonance in Medicine
Volume38
Issue number6
DOIs
Publication statusPublished - Dec 1997
Externally publishedYes

Keywords

  • Magnetic resonance image encoding
  • Non-Fourier encoded MRI
  • Spatially selective RF excitation

Fingerprint

Dive into the research topics of 'Non-Fourier encoding with multiple spin echoes'. Together they form a unique fingerprint.

Cite this