Abstract
The proximal part of the radius has a complex shape and dimension that cannot be precisely determined by standard roentgenogram for real three-dimensional anatomical shape which is important for prosthesis design. This study presents a method by which computer tomography (CT) images are combined with the reverse engineering technique to obtain and analyse the three-dimensional inner and outer geometry of the proximal radius. The three-dimensional models were reconstructed from CT images obtained from 40 radial bones and approximated with two- and three-dimensional fitting algorithms based on reverse engineering methods. The mean total length of the radius was 240.0 mm [standard deviation (SD) = 17.3]. The radial head in this study is more likely to be circular with an average diameter of 20.5 mm (SD = 1.9). The outer diameter of the radial neck averages 14.7 mm (SD = 1.0). The thickness of the radial head averages 12.9 mm (SD = 1.4). The intramedullary canal diameter of the radial neck averages 7.4 mm (SD = 1.4). The depth of the fossa at the articular surface averages 1.5 mm (SD = 0.4).
| Original language | English |
|---|---|
| Pages (from-to) | 79-84 |
| Number of pages | 6 |
| Journal | Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine |
| Volume | 218 |
| Issue number | 1 |
| DOIs | |
| Publication status | Published - 2004 |
Keywords
- Anatomic data
- Radial head prosthesis
- Reversed engineering
Fingerprint
Dive into the research topics of 'Morphology of the radial head: A reverse engineering based evaluation using three-dimensional anatomical data of radial bone'. Together they form a unique fingerprint.Cite this
- APA
- Author
- BIBTEX
- Harvard
- Standard
- RIS
- Vancouver