Influence of an object's z-axis location and location on the axial plane on the voxel value representation and uniformity in cone beam computed tomography

  • Kanako Nishino
  • , Hiroaki Shimamoto
  • , Naoya Kakimoto
  • , Tomomi Tsujimoto
  • , Jira Chindasombatjaroen
  • , Shumei Murakami
  • , Souhei Furukawa
  • Osaka University

Research output: Contribution to journal › Article › peer-review

3 Citations (Scopus)

Abstract

Objective This study aimed to determine the influence of an object's z-axis location and location on the axial plane on the voxel values in cone beam computed tomography.

Study Design The CTP401 and CTP486 of the Catphan 500 phantom were scanned with an Alphard-3030 at 6 locations in the z-axis.

Results In CTP401, the voxel values of 3 inserts showed the lowest values when the center of the radiation field corresponded to under 2 or 4 cm of the z-axis center of CTP401. In CTP486, the voxel values showed the lowest values when the center of the radiation field corresponded to under 2 or 4 cm of the z-axis center of CTP486. On the same axial plane, the voxel values at the center showed significantly lower values than the off-center locations (P =.002).

Conclusions The voxel values in the Alphard-3030 changed based on the z-axis location of the radiation field and were nonuniform on the same axial plane.

Original languageEnglish
Pages (from-to)619-624
Number of pages6
JournalOral Surgery, Oral Medicine, Oral Pathology and Oral Radiology
Volume118
Issue number5
DOIs
Publication statusPublished - 1 Nov 2014

Fingerprint

Dive into the research topics of 'Influence of an object's z-axis location and location on the axial plane on the voxel value representation and uniformity in cone beam computed tomography'. Together they form a unique fingerprint.

Cite this