Implementing log file-based patient-specific QA for VMAT plans: A comparative study of MobiusFX and measurement-based approaches

  • Faculty of Medicine Ramathibodi Hospital, Mahidol University

Research output: Contribution to journal › Article › peer-review

1 Citation (Scopus)

Abstract

Background: The current interest in utilizing log file-based patient-specific quality assurance (PSQA) is growing as an alternative to measurement-based approaches for volumetric modulated arc therapy (VMAT) plans. However, implementing log file-based PSQA requires further experience and understanding since there are no guidelines. Purpose: This study aimed to evaluate the feasibility of implementing commercial log file-based PSQA (MobiusFX) for pretreatment verification of VMAT plans by comparing its outcomes with measurement-based methods. Methods: The study was conducted by selecting 40 VMAT plans and these plans were performed for pretreatment verification for both MobiusFX and measurement-based methods, including portal dosimetry, ArcCHECK, and Octavius4D with gamma analysis at criteria 3%/3 mm, 3%/2 mm, and 2%/2 mm. The agreement of gamma passing rate between MobiusFX and measurement-based methods was assessed through scatter plots and percentage agreement score. Additionally, the root mean square (RMS) errors obtained from MobiusFX were investigated in each machine parameter, and examined the correlation between the gamma passing rate and plan complexity metric (MU/Gy) by Spearman's rank correlation coefficient. The statistical control process concept from TG 218 was also applied to establish the action limit (AL) and tolerance limit (TL) for MobiusFX. Results: The scatter plots of MobiusFX illustrated a trend similar to the measurement-based methods, with high percentage agreement scores observed at 3%/3 mm and 3%/2 mm criteria. RMS errors were zero for all parameters except multileaf collimator (MLC), and the maximum RMS was 0.06 mm. No correlation was found between the gamma passing rate and the complexity metric in this study. The most appropriate TL and AL for the gamma passing rate with MobiusFX were determined to be 95.38% and 94.75%, respectively, at the 3%/2 mm criterion. Conclusion: MobiusFX showed interchangeable gamma passing rates with measurement-based methods at the 3%/3 mm and 3%/2 mm criteria, supporting its use as an alternative for VMAT pretreatment verification. Machine- and system-specific factors should be considered for reliable clinical implementation.

Original languageEnglish
Article numbere70526
JournalJournal of Applied Clinical Medical Physics
Volume27
Issue number3
DOIs
Publication statusPublished - Mar 2026

Keywords

  • MobiusFX
  • VMAT
  • log file-based PSQA
  • patient-specific QA

Fingerprint

Dive into the research topics of 'Implementing log file-based patient-specific QA for VMAT plans: A comparative study of MobiusFX and measurement-based approaches'. Together they form a unique fingerprint.

Cite this