Finite element study of the proximal femur with retained trochanteric gamma nail and after removal of nail

  • B. Mahaisavariya
  • , K. Sitthiseripratip
  • , J. Suwanprateeb
  • Thailand Science Park

Research output: Contribution to journal › Article › peer-review

40 Citations (Scopus)

Abstract

This study aims to evaluate the stress and strain distributions in the healed proximal femur after fixation with a trochanteric gamma nail (TGN) and after TGN removal, using the finite element method. The stress distributions in the proximal femur with retained TGN and after TGN removal were very similar. The strain and the strain energy density in the femoral neck region with retained TGN were much higher than in the lag screw hole at the subtrochanter and the distal locking screw hole at the proximal femur, and even higher after TGN removal. Stair climbing resulted in higher strain and higher strain energy density at the femoral neck than normal walking. The conclusion can be drawn that removal of the TGN may result in high risk of femoral neck fracture.

Original languageEnglish
Pages (from-to)778-785
Number of pages8
JournalInjury
Volume37
Issue number8
DOIs
Publication statusPublished - Aug 2006

Keywords

  • Finite element study
  • Fracture
  • Gamma nail
  • Intramedullary nail
  • Trochanteric fracture

Fingerprint

Dive into the research topics of 'Finite element study of the proximal femur with retained trochanteric gamma nail and after removal of nail'. Together they form a unique fingerprint.

Cite this