Evaluation of the Bedside Glucose Monitoring System in Neonatal Units

  • Maharat Nakhonratchasima Hospital
  • Ramathibodi Hospital

Research output: Contribution to journal › Article › peer-review

1 Citation (Scopus)

Abstract

Objective: To evaluate the performance of a commercial® reader compared with a laboratory method in neonates with a variety of diseases and conditions. Patients and Method: A total of 175 patients were included in the present study. Venous whole blood samples were analyzed with a commercial® reader by trained nurse. Through the same sampling site, specimens were collected, spun and plasma were sent to the laboratory for measurement of plasma glucose. Results: The regression analysis between the results of a commercial® reader and laboratory glucose were significantly correlated (r = 0.97; p < 0.001) with the result as follows: A commercial® reader = Laboratory glucose - 0.17 (n = 175). A positive slope of 0.04 was found between hematocrit and difference between a commercial® reader and laboratory plasma glucose. However, this correlation was of little clinical significance. Conclusions: A commercial® reader showed a good correlation with the standard laboratory method for the measurement of plasma glucose in neonates with a variety of diseases and conditions.

Original languageEnglish
Pages (from-to)883-888
Number of pages6
JournalJournal of the Medical Association of Thailand
Volume86
Issue number9
Publication statusPublished - Sept 2003

Keywords

  • A Commercial® Reader
  • Blood Glucose
  • Hyperglycemia
  • Hypoglycemia
  • Neonatal

Fingerprint

Dive into the research topics of 'Evaluation of the Bedside Glucose Monitoring System in Neonatal Units'. Together they form a unique fingerprint.

Cite this