Endotracheal tubes cleaned with a novel mechanism for secretion removal: A randomized controlled clinical study

  • Riccardo Pinciroli
  • , Cristina Mietto
  • , Annop Piriyapatsom
  • , Christopher T. Chenelle
  • , John G. Thomas
  • , Massimiliano Pirrone
  • , Lynn Bry
  • , Gregory R. Wojtkiewicz
  • , Matthias P. Nahrendorf
  • , Robert M. Kacmarek
  • , Lorenzo Berra
  • Massachusetts General Hospital
  • West Virginia University School of Dentistry
  • Brigham and Women's Hospital

Research output: Contribution to journalArticlepeer-review

27 Citations (Scopus)

Abstract

INTRODUCTION: Intubation compromises mucus clearance, allowing secretions to accumulate inside the endotracheal tube (ETT). The purpose of this trial was to evaluate a novel device for ETT cleaning. We hypothesized that its routine use would reduce tube occlusion due to mucus accumulation, while decreasing airway bacterial colonization. METHODS: Subjects were randomized to either the use of the device every 8 h, or the institutional standard of care (blind tracheal suction) only. ETTs were collected at extubation and analyzed with high-resolution computed tomography (HRCT) for quantification of mucus volume. Microbiological testing was performed on biofilm samples. Vital signs and ventilatory settings were collected at the bedside. In-hospital follow-up was conducted, and a final evaluation survey was completed by respiratory therapists. RESULTS: Seventy-four subjects expected to remain intubated for longer than 48 h were enrolled (77 ETTs, 37 treatment vs 40 controls). Treated tubes showed reduced mucus accumulation (0.56 ± 0.12 vs 0.71 ± 0.28 mL; P = .004) and reduced occlusion (6.3 ± 1.7 vs 8.9 ± 7.6%; P = .039). The HRCT slice showing the narrowest lumen within each ETT exhibited less occlusion in cleaned tubes (10.6 ± 8.0 vs 17.7 ± 13.4%, 95% CI: 2–12.1; P = .007). Data on microbial colonization showed a trend in the treatment group toward a reduced ETT-based biomass of bacteria known to cause ventilator-associated pneumonia. No adverse events were reported. The staff was satisfied by the overall safety and feasibility of the device. CONCLUSION: The endOclear is a safe and effective device. It prevents luminal occlusion, thereby better preserving ETT nominal function.

Original languageEnglish
Pages (from-to)1431-1439
Number of pages9
JournalRespiratory Care
Volume61
Issue number11
DOIs
Publication statusPublished - 1 Nov 2016
Externally publishedYes

UN SDGs

This output contributes to the following UN Sustainable Development Goals (SDGs)

  1. SDG 3 - Good Health and Well-being
    SDG 3 Good Health and Well-being

Keywords

  • Airway obstruction
  • Biofilms
  • Intubation
  • VAP (Ventilator-associated pneumonia)
  • Ventilation

Fingerprint

Dive into the research topics of 'Endotracheal tubes cleaned with a novel mechanism for secretion removal: A randomized controlled clinical study'. Together they form a unique fingerprint.

Cite this