Skip to main navigation Skip to search Skip to main content

Affordable technology for enumeration of the absolute CD4 T-lymphocyte count by cell bead assay

  • Chulalongkorn University
  • University of Tsukuba

Research output: Contribution to journal › Article › peer-review

3 Citations (Scopus)

Abstract

The quantitative BD Trucount (San Jose, CA) tube method is the conventional but expensive method to quantitate CD4+T-lymphocyte (CD4) counts, and this may be beyond the means of countries with limited resources. In this study, we validated a quantitative method known as a cell-bead (CB) assay to quantitate CD4 counts in the peripheral blood of healthy subjects. The absolute CD4 count obtained from the CB method was highly correlated with those obtained from the Trucount tube (r2=0.98, y=26.73+1.01x, P<0.0001 and a mean bias of 34.8 cell/μL, limit of agreement [LOA] -34.8-104.4 cell/μL) and flow rate-based assay method (r2=0.97; y=69.51 + 0.88x, P<0.0001 and a mean bias -53.5 cell/μL, LOA -149.4-42.3 cell/μL). This study demonstrates that the CB method is suitable and more affordable for CD4 quantitation. This method is inexpensive and interchangeable with the latex bead-based methods for generating absolute counts in resource-limited areas.

Original languageEnglish
Pages (from-to)423-428
Number of pages6
JournalLab Medicine
Volume41
Issue number7
DOIs
Publication statusPublished - Jul 2010

Keywords

  • Absolute count
  • Cell bead
  • Flow cytometry
  • Flow rate based calibration
  • Trucount

Fingerprint

Dive into the research topics of 'Affordable technology for enumeration of the absolute CD4 T-lymphocyte count by cell bead assay'. Together they form a unique fingerprint.

Cite this