Abstract
A 58-year-old man with an ischemic cardiomyopathy underwent ICD implantation after having an electrophysiological study positive for inducible, sustained monomorphic VT. Approximately 1 week later he experienced a shock. Inspection of stored electrograms revealed that a run of nonsustained VT had initiated his ICD to charge. The episode spontaneously terminated, yet the shock was still delivered during sinus rhythm. Further inspection revealed that a critically timed PVC during the "second look" interval resulted in delivered therapy. Thus, inappropriate ICD therapy can result from causes other than SVT, lead malfunction, or over sensing.
| Original language | English |
|---|---|
| Pages (from-to) | 1403-1405 |
| Number of pages | 3 |
| Journal | PACE - Pacing and Clinical Electrophysiology |
| Volume | 25 |
| Issue number | 9 |
| DOIs | |
| Publication status | Published - 1 Sept 2002 |
| Externally published | Yes |